Test Code PROG PROGESTERONE, SERUM
Additional Codes
Software | Test Code |
---|---|
Label Text | PROG |
EPIC | LAB529 |
Performing Laboratory
NorDx Laboratories
Useful For
Diagnosing and treating disorders of the ovaries or placenta; detecting if ovulation has occurred; monitoring some patients with adrenal and testicular tumors
Method Name
Electrochemiluminescent Immunoassay (ECLIA)
Reference Values
Healthy Females
Follicular: <0.05 - 0.193 ng/mL
Ovulation: 0.055 - 4.14 ng/mL
Luteal: 4.11 - 14.5 ng/mL
Post-menopause: <0.05 - 0.13 ng/mL
Healthy Pregnant Females
1st trimester: 11.0 - 44.3 ng/mL
2nd trimester: 25.4 - 83.4 ng/mL
3rd trimester: 58.7 - 214 ng/mL
Males: <0.05 - 0.15 ng/mL
Days and Times Test Performed
Monday through Sunday
Report Available
1 Day
Available Stat
Specimen Type
Serum
Preferred Container
Serum Separator Tube (SST)
Preferred Volume
Serum: 1 mL (Serum Separator Tube (SST))
Minimum Volume
Serum: 0.5 mL (Serum Separator Tube (SST))
Specimen Collection and Handling
1. Spin specimen, separate from clot and send refrigerated.
2. Sample should not be collected on patients receiving therapy with high biotin doses (i.e. >5 mg/day) until at least 8 hours after the last biotin administration.
Specimen Stability Information
Specimen Type: Serum
Frozen: 3 Months
Refrigerated: 7 Days
Must be spun/separated within: 2 Hours
Add On Capable
Yes
Advance Beneficiary Notice Requirements
No ABN Required
CPT Code Information
CPT Code |
CPT Description |
CPT Disclaimer |
---|---|---|
84144 | Progesterone, Quantitative |
Performing Laboratory Location
NorDx Laboratories
Clinical Significance
No sample should be collected on patients receiving therapy with high biotin doses (i.e. biotin therapy for Multiple Sclerosis or oncology patients; skin, hair and nail supplements, or multivitamins containing > 5 mg/day) until at least 12 hours after the last biotin administration. If unsure, or if the clinical picture does not fit the results please contact the laboratory . We have methods to check for biotin interference as well as for any other interferences and for the accuracy of the results.
Biotin interference would falsely increase the result of this assay.
LOINC Code Information
2839-9
Acceptable Alternative Container(s)
Plasma Separator Tube (PST)
Red Top Tube