Test Code INHIX FACTOR IX (9) INHIBITOR WITH ACTIVITY
Additional Codes
Software | Test Code |
---|---|
Label Text | INHIX |
EPIC | LAB20290 |
Performing Laboratory
NorDx Laboratories
Useful For
Detecting the presence and titer of a factor inhibitor directed against Human Factor IX
Method Name
Automated Clot-Based Assay
Reference Values
<0.6 BIU/mL
Days and Times Test Performed
Varies
Report Available
Approximately 48 hours for STAT collections
Approsimately 1 week for routine collections
Specimen Type
Plasma
Preferred Container
2 x Light Blue Top Tube (Sodium Citrate) – Tubes must be full at time of collection.
Specimen Collection and Handling
Collect specimen as follows:
1. Draw blood in light blue-top (3.2% sodium citrate) tubes. Tubes must be full upon collection.
2. Separate plasma from cells as soon as possible within 4 hours of collection in a centrifuge at 2,500 rpm for 15 minutes or 3,000 rpm for 10 minutes.
3. Using a plastic transfer pipette (do not use a glass pipette) decant the plasma to a screw-capped, plastic (polypropylene) vial. Glass vials will not be accepted.
4. Spin the plasma again at 2,500 rpm for 15 minutes or 3,000 rpm for 10 minutes. Note: Double spinning the specimen ensures platelet poor plasma (<10,000 uL)
5. Divide the plasma into as many 0.5-1.0 mL aliquots as possible and cap the vial(s) tightly.
6. Place tubes upright and flash freeze specimen(s) immediately at ≤ – 20° C or below.
7. Send specimen(s) FROZEN on dry ice.
Click here to view the Procedure for Collecting a Specimen from an A-line
Specimen Stability Information
Specimen Type: Plasma
Room Temp: 4 Hours
Note: Whole blood that remains capped and unspun is stable for up to 4 hours when stored at room temperature (18 – 24C). Frozen plasma is stable 2 weeks @ -20C and 12 months @ -70C. Blue top tubes for factor assays must not be refrigerated or placed on ice.
Add On Capable
Not Permitted
CPT Code Information
CPT Code |
CPT Description |
CPT Disclaimer |
---|---|---|
85250 | Clotting, Factor IX (PTC or Christmas) | |
85335 | Factor Inhibitor Test |
Performing Laboratory Location
NorDx Laboratories
Rejection Information
Specimens will be rejected if underfilled (<90 %), clotted, high hematocrit (>55 %), or grossly hemolyzed