Sign in →

Test Code DOABS / SDS1A-MC DRUGS OF ABUSE, SERUM

Additional Codes

Software Test Code
SoftID DOABS
EPIC LAB20320
Michigan SDS1A

Performing Laboratory

Michigan Co-Tenancy Laboratory at Warde Medical Laboratory (MCL)

Useful For

Although urine is the specimen of choice for drug screening, a serum drug screen can be performed when a urine specimen is unavailable.

Method Name

Immunoassay (IA)

Reference Values

Decision Levels:

Drugs Included Screen Cutoff Concentration

Amphetamines 500 ng/mL

Barbiturates 150 ng/mL

Benzodiazepines 100 ng/mL

Cocaine 150 ng/mL

Ethanol 10 mg/dL*

Methadone 150 ng/mL

Opiates 150 ng/mL

Phencyclidine 12 ng/mL

Propoxyphene 150 ng/mL

THC (Cannabis) 50 ng/mL

*Quantitative value reported.

Days and Times Test Performed

Monday through Friday

Report Available

3 Days

Specimen Type

Serum

Preferred Container

Red Top Tube

Preferred Volume

Serum: 1 mL (Red Top Tube)

Minimum Volume

Serum: 0.5 mL (Red Top Tube)

Specimen Collection and Handling

Spin specimen, separate from clot and send refrigerated in a screw-capped, plastic vial.

Specimen Stability Information

Specimen Type: Serum

Frozen: 14 Days

Refrigerated: 14 Days

Room Temp: 48 Hours

Add On Capable

Contact Reference Lab – If add-on can never be done, report the test to the Administrative Analyst

Performing Laboratory Location

Michigan Co-Tenancy Laboratory at Warde Medical Laboratory (MCL)

Acceptable Alternative Container(s)

Lavender Top Tube (EDTA) – EDTA Plasma acceptable (Citrate and Oxalate plasma also acceptable, though not preferred)

Rejection Information

Sample collected in an Serum separator tube (SST), gross hemolysis, lipemia, any tubes containing gel