603599Useful For
Investigating the differential diagnosis for patients with water balance disorders, including diabetes insipidus, in conjunction with osmolality and hydration status
May aid in the evaluation of cardiovascular disease in conjunction with other cardiac markers
Clinical Information
Arginine vasopressin (AVP) and copeptin (also known as copeptin proAVP or copeptin AVP) are derived from the same precursor peptide. Copeptin has been proposed as a more stable, potentially superior, surrogate marker of AVP in the assessment of water balance disorders. Unlike AVP, copeptin is stable in plasma. Both copeptin and AVP are responsive to osmotic stimuli and increase in response to water deprivation. In healthy subjects, water deprivation causes the plasma osmolality to rise above approximately 280 to 290 mOsmol/kg, leading to the release of AVP and copeptin into the circulation. Copeptin increases gradually with fasting and water deprivation and declines rapidly after intake of water and food.
Diabetes insipidus (DI) is characterized by the inability to appropriately concentrate urine in response to volume and osmolar stimuli. The main causes for DI are decreased AVP production (central DI) or decreased renal response to AVP (nephrogenic DI).
The determination of the underlying disease pathology in patients with polyuria and altered plasma osmolality is often difficult. Polyuria can be related to insufficient AVP (central DI), reduced sensitivity to AVP (nephrogenic DI), or excessive water intake. Measurement of plasma copeptin concentration has been shown to be useful in the investigation of these AVP-related disorders. Additionally, utilization of copeptin has been proposed in the assessment of syndrome of inappropriate antidiuretic.
Copeptin is also a marker of acute hemodynamic stress and has been reported to aid in the prognosis or diagnosis of several cardiac disorders, such as acute coronary syndrome, stable coronary artery disease, congestive heart failure, and acute ischemic stroke. Some studies have demonstrated that copeptin may improve prediction of mortality and heart disease outcome when combined with natriuretic peptides such as B-type natriuretic peptide (BNP) and N-terminal proBNP.
Method Name
Immunofluorescent Assay (IFA)
Reporting Name
Copeptin proAVP, PReference Values
Non-water deprived, non-fasting adults*: <13.1 pmol/L
Water deprived, fasting adults**: <15.2 pmol/L
Non-water deprived, non-fasting pediatric patients***: <14.5 pmol/L
Note:
*Keller T, Tzikas S, Zeller T, et al. Copeptin improves early diagnosis of acute myocardial infarction. J Am Coll Cardiol. 2010;55(19):2096-2106. doi:10.1016/j.jacc.2010.01.029
**Internal Mayo Clinic study
***Du JM, Sang G, Jiang CM, He XJ, Han Y. Relationship between plasma copeptin levels and complications of community-acquired pneumonia in preschool children. Peptides. 2013;45:61-65. doi:10.1016/j.peptides.2013.04.015
Specimen Type
Plasma EDTASpecimen Required
Patient Preparation: For water-deprivation testing, for at least 8 hours, the patient should fast and thirst (no liquids, including water, are allowed).
Collection Container/Tube: Lavender top (EDTA)
Submission Container/Tube: Plastic screw-top vial
Specimen Volume: 0.5 mL
Collection Information: Centrifuge and aliquot plasma into a plastic vial. Do not submit in original tube.
Specimen Minimum Volume
0.3 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Plasma EDTA | Refrigerated (preferred) | 7 days |
Frozen | 30 days | |
Ambient | 7 days |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Cautions
Sepsis, severe sepsis, septic shock, lower respiratory tract infections, chronic obstructive pulmonary disease, and cardiovascular diseases (eg, chronic heart failure) may increase copeptin concentrations.
Arginine vasopressin (AVP) receptor antagonist therapies and other diseases in which AVP has been shown to play an important pathophysiologic role may also increase copeptin concentration.
In some cases, bronchial carcinoma may lead to ectopic copeptin secretion.
Mixed forms of diabetes insipidus (DI) can exist, and both central and peripheral DI may be incomplete, complicating the interpretation of results.
In rare cases, some individuals can develop antibodies to mouse or other animal antibodies (often referred to as human anti-mouse antibodies [HAMA] or heterophile antibodies), which may cause interference in some immunoassays. Caution should be used in interpretation of results and the laboratory should be alerted if the result does not correlate with the clinical presentation.
Day(s) Performed
Monday through Saturday
Report Available
Same day/1 to 3 daysPerforming Laboratory

Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
84588
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
CPAVP | Copeptin proAVP, P | 78987-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
CPAVP | Copeptin proAVP, P | 78987-5 |
Method Description
Copeptin pro-arginine vasopressin (proAVP) is measured in this homogeneous automated immunofluorescent assay on the BRAHMS Kryptor Compact PLUS. The Kryptor Compact PLUS uses TRACE (time resolved amplified cryptate emission) technology based on a nonradioactive transfer of energy. This transfer occurs between 2 fluorescent tracers: the donor (europium cryptate) and the acceptor (XL707). In the Copeptin proAVP assay, a sheep polyclonal antibody against C-terminal proAVP is labeled with europium cryptate and a mouse monoclonal antibody against C-terminal proAVP is labeled with XL707. Copeptin is sandwiched between the 2 antibodies, bringing them into close proximity. When the antigen-antibody complex is excited with a nitrogen laser at 337 nm, some fluorescent energy is emitted at 620 nm and the rest is transferred to XL707. This energy is then emitted as fluorescence at 707 nm. A ratio of the energy emitted at 707 nm to that emitted at 620 nm (internal reference) is calculated for each sample. Signal intensity is proportional to the number of antigen-antibody complexes formed and, therefore, to antigen concentration.(Unpublished Mayo method)